    Electron Spin Resonance Model No. esr

    Compact ESR(EPR) Systems for Biodosimetry

    Model No. esr34


    The newly-developed spectrometer is a lightweight (27 kg) one-box ESR device with a resonator showing a Q-factor higher than that of a previous Keycom model, which is quieter, without influence from magnetic modulation, and contains a fingernail positioner.
    The best measurement result obtained after the cavity Q-factor was increased to more than 7,200 by continuous polishing of the inner surface of the cavity using deerskin.
    The common mode noise of “magic T” was also successfully decreased to as low as one-half by completely tuning the arm balance.
    These efforts markedly decreased the noise level and extended downwardly the linear portion of dose dependence as low as 0.4 Gy.

    Best Suitable for Evaluation Measurement of Accumulated Absorbed Dose Published Related Paper

    • Development of a compact electron spin resonance system for measuring esr signals of irradiated fingernails
      Suzuki, Hirosuke*; Tamukai, Kenji*; Yoshida, Naoki*; Ohya, Hiroaki*; Kato, Katsuhisa*; Anzai, Kazunori†; Swartz, Harold M.‡

    Method of Evaluation


    • An ESR signal (left figure) depends on irradiation dose (left).
      Its magnitude can simply estimated by the difference (height) between maximum and minimum of the signals of human nails.
      Thus, the potential benefits of ESR dosimetry using human nails include the ready availability of samples for dose measurements and the capability to make immediate dose assessment.
      Fingernails consist mainly of a-keratin, and the measured ESR signals appear to originate from radiation-induced radicals formed in this component [i.e., radiation-induced signals (RIS)] as well as from mechanically-induced radicals formed by clipping nails for sampling [i.e., mechanically-induced signals (MIS)].
      This mechanical signal appears without irradiation.
      However, MIS concentrates around cutting surfaces, clipping nails dipped into water and wipes out by paper towel.
      This procedure can remarkably diminish MIS contribution.

    Measurement Process

    • Connect on-hand PC to ESRX-10SA-v4,and power them on. 「Adjusting Window」 will automatically appear on the PC, and measuring process will begin.

      • By「Resonator Adjust Window]resonator characteristics are adjusted, and after setting measuring conditions at 「Condition Setting Window」, all are controlled through [Measurement Window」.
      • According to the object of the measurements, some integration, accumulation, adjustment of blank data etc. moving average can be selected, as required.
      • Data processing such as peak height(h), first and second Integration are selected as required. ・This system can automatically correct the magnetic field by using the position of the ESR signals of Mn2+ (option)

    Model No.

    • ESRX-10SA-v4
      * Windows PC is not included.

    Specification