Improved Production Efficiency Reduces Measurement and Inspection Time The MS9740A is a benchtop optical spectrum analyzer covering 600 nm to 1750 nm. The MS9740A offers high-performance optical resolution and high-speed measurements to support the needs of device and component manufacturers. Waveform data can be measured and transmitted to external control equipment within 0.2s -- a 5x improvement over previous Anritsu models. This enables optical device manufacturers to reduce costs and time for testing and analysis by:
- Improving sweep speed (measurement time), and
- Accelerating data transfer to control equipment after measurement Combine these overall improvements with the ability to do basic analysis, and the result is excellent return on investment and improved productivity.
|
Specificati>Details |
Applicable optical fiber |
10 µm/ 125 µm SM fiber (ITU-T G.652), 50 µm/ 125 µm GI fiber |
Optical connector |
User replaceable: FC, SC, ST, DIN, LC |
Wavelength measurement range |
600 nm to 1750 nm |
Wavelength accuracy *1 |
±20 pm (1520 nm to 1620 nm, Resolution: 0.03 to 0.2 nm) *2,
±100 pm (1520 nm to 1620 nm, Resolution: 0.5, 1.0 nm) *2 |
Setting resolution |
0.03, 0.05, 0.07, 0.1, 0.2, 0.5, 1.0 nm
(RBW: 3dB optical filter: transmission bandwidth) |
Maximum input power |
+23 dBm (Optical ATT: On) |
Dynamic range *1 |
High dynamic range: 60 dB (0.4 nm from peak wavelength)
Normal dynamic range: 58 dB (0.4 nm from peak wavelength)
[Resolution: 0.05 nm, 1550 nm, Optical Att: Off, 20 to 30°C] |
Sweep |
Sweep width: 0.2 nm to 1200 nm, 0 nm
Sweep speed: ≤0.2 sec. (span 5 nm)
[Resolution: 0.1 nm, VBW=10 kHz, Normal dynamic range, center 1550 nm, sweep start to stop, no optical input, sampling point: ≤501] |
Dimensions and mass |
426 (W) x 177 (H) x 350 (D) mm (excluding projections) ≤15.0 kg (without option) |
|
*1: Using SM fiber (ITU-T G.652), after 2 hours of warm-up (the Repeat sweeping performed at Span 100 nm or more and VBW 10 kHz or more during the warm-up operation), after Auto Align, at stable room temperature.
*2: Built-in option 002, after Wl cal (ref) wavelength calibration execution, at stable room temperature. |
|