Specifications |
Resistance |
Nominal 100 ohms at 0°C |
Temperature Coefficient |
0.00385 ohms/ohm/°C nominal |
Temperature Range |
5627-12: –200°C to 420°C
5627-9: –200°C to 420°C
5627-6: –200°C to 300°C
(transition and cable temperature 150°C maximum) |
Resistance Stability |
±0.13°C at 0°C after 1000 hours at 400°C |
Sheath Material |
316 Stainless Steel |
Leads |
Teflon™-insulated, nickel-plated stranded copper, 22 AWG |
Termination |
Four gold-plated spade lugs are standard. Specify 1502 connector if needed. Other options available. |
Hysteresis |
0.04% maximum between –200°C and 420°C |
Time Constant |
Four seconds maximum for 63.2% response to step change in water moving at 3 fps. |
Bending Radius |
Sheath is bendable on a minimum radius of 3/4 inch except for 2-inch sensitive area of sheath near tip. |
Calibration |
Includes NIST-traceable table with R vs. T values in 1°C increments from –183°C to 500°C. ITS-90 coefficients included. |
Probe Accuracy (includes calibration uncertainty and short-term stability) |
–196°C ±0.030°C
0°C ±0.050°C
200°C ±0.051°C
420°C ±0.055°C |
Size |
5627-12: 12" L x 1/4" Dia.
5627-9: 9" L x 3/16" Dia.
5627-6: 6" L x 3/16" Dia. |