Specifications |
6020 |
6022 |
6024 |
Range |
20°C to 300°C† |
Stability |
±0.001°C at 40°C (water)
±0.003°C at 100°C (oil)
±0.005°C at 300°C (oil) |
Uniformity |
±0.002°C at 40°C (water)
±0.004°C at 100°C (oil)
±0.012°C at 300°C (oil) |
Temperature Setting |
Digital display with push-button data entry |
Set-Point Resolution |
0.01°C; high-resolution mode, 0.00018°C |
Display Temperature Resolution |
0.01°C |
Digital Setting Accuracy |
±1°C |
Digital Setting Repeatability |
±0.02°C |
Heaters |
350 and 1050 watts |
Access Opening
(call for custom openings) |
5" x 10"
(127 x 254 mm) |
5" x 10"
(127 x 254 mm) |
7.25" x 12.75"
(184 x 324 mm) |
Depth |
12" (305 mm) |
18.25" (464 mm) |
13.25" (337 mm) |
Wetted Parts |
304 stainless steel |
Power |
115 VAC (±10%), 50/60 Hz, 10 A or 230 VAC (±10%), 50/60 Hz, 5 A, specify, 1075 W |
Volume |
7.2 gallons
(27 liters) |
11.1 gallons
(42 liters) |
11.2 gallons
(42 liters) |
Weight |
70 lb. (32 kg) |
80 lb. (36 kg) |
80 lb. (36 kg) |
Size |
25.5" H x 16" W x 20" D
(648 x 406 x 508 mm) |
32" H x 16" W x 20" D
(813 x 406 x 508 mm) |
27.5" H x 19" W x 23" D
(699 x 483 x 584 mm) |
Automation Package |
Interface-it software and RS-232 computer interface are available for setting bath temperature via remote computer. For IEEE-488, add the 2001-IEEE to the automation package. |
†External cooling required for operation below 40°C. Cooling coils are built into the bath walls. Tubing ports are accessible at the back of the bath for circulating chilled fluid or shop air to boost cooling. |