Specifications |
Range |
300°C to 1100°C
(572°F to 2012°F) |
Stability |
±0.05°C at 300°C
±0.1°C at 700°C
±0.1°C at 1100°C |
Uniformity |
±0.05°C at 300°C
±0.2°C at 700°C
±0.3°C at 1100°C |
Heating Rates |
700°C to 900°C, 145°C/hour on low; 900°C to 1000°C, 370°C/hour on high |
Cooling Rates |
Nom. at 800°C ³300°C/hour Nom. at 600°C ³180°C/hour |
Stabilization Time |
Typically 2 hours midrange, slower at low-temperature end (4 hours), faster at high-temperature end |
Interface |
RS-232 included on all units |
Outside Dimensions |
18" H x 13.25" W x 26" D (457 x 337 x 660 mm) |
Thermal Blocks |
Block A: 11.5" (292 mm) immersion
Block B: 16" (406 mm) immersion (for Au-Pt TCs)
Each includes four wells at 1/4" (6.35 mm) |
Weight |
72.5 pounds (33 kg) complete [63.5 pounds (28.8 kg) w/o block] |
Power |
230 VAC (±10%), 50/60 Hz,
16 A, 3700 W |
Heater |
3700 W high,
925 W low |
NIST-Traceable Calibration |
Data at 420°C |