Specifications |
1575 |
1590 |
|
Nominal
Resistance
|
Accuracy (of indicated value) |
Equivalent Temperature
Value, at 0°C
|
Nominal
Resistance
|
Accuracy (of indicated value) |
Equivalent Temperature
Value, at 0°C
|
Transfer Accuracy (using external reference resistor) |
0.25W
2.5W
25W
100W
10 KW |
40 ppm
20 ppm
4 ppm
4 ppm
10 ppm |
0.01°C
0.005°C
0.001°C
0.001°C
0.00025°C (thermistor at 25°C) |
0.25W
2.5W
25W
100W
10 KW |
20 ppm
5 ppm
1 ppm
1 ppm
5 ppm |
0.005°C
0.00125°C
0.00025°C
0.00025°C
0.000125°C (thermistor at 25°C) |
Absolute Accuracy (using internal reference resistor) |
0.25W
2.5W
25W
100W
10 KW |
100 ppm
40 ppm
8 ppm
8 ppm
20 ppm |
0.025°C
0.01°C
0.002°C
0.002°C
0.0005°C (thermistor at 25°C) |
0.25W
2.5W
25W
100W
10 KW |
40 ppm
20 ppm
6 ppm
6 ppm
10 ppm |
0.01°C
0.005°C
0.0015°C
0.0015°C
0.00025°C (thermistor at 25°C) |
Typical Resolution |
0.25W
2.5W
25W
100W
10 KW |
10 ppm
5 ppm
1 ppm
1 ppm
3 ppm |
0.0025°C
0.00125°C
0.00025°C
0.00025°C
0.000075°C (thermistor at 25°C) |
0.25W
2.5W
25W
100W
10 KW |
10 ppm
2 ppm
0.5 ppm
0.5 ppm
2 ppm |
0.0025°C
0.0005°C
0.000125°C
0.000125°C
0.00005°C (thermistor at 25°C) |
Resistance Range |
0W to 500 KW |
Internal Reference Resistors |
1W, 10W, 100W, 10 KW |
Minimum Measurement Period |
2 seconds |
Current Source |
0.001 mA to 15 mA, programmable |
Analog Output |
–5 to +5 V |