These liquid separators provide a barrier between the deadweight tester operating fluid and that of the device under test.
These instruments connect directly to the test port of a hydraulic deadweight tester and by means of a flexible diaphragm separate the fluids thus protecting the calibrator from contamination.
*5521-For pressures to 10,000 psi (700 bar)
*5522-For pressures to 7,000 psi (480 bar) for Skydrol or other aggressive fluids
Specifications |
Pressure Range |
5521: 0 to 10,000 psi (700 bar)
5522: 0 to 7,000 psi (480 bar) for Skydrol and other aggressive fluids |
Instrument Size (W x H) |
4 x 6 in (100 x 150 mm) |
Instrument Weight |
7.5 lb (3.5 kg) |
Test Port Adapter |
Uses adapters supplied with calibrator |
Calibrator Connection |
3/8 BSP male (mating adapter supplied) |
Body Material |
Aluminum bronze |
Diaphragm |
5521- Viton® 5522-Ethylene Propylene |
Reservoir Material |
Hard-anodized aluminum |
|