7350 Pneumatic high pressure controller |
|
|
|
|
The Model 7350 High Pressure Pneumatic Controller and Model 7615 High Pressure Hydraulic Controller provide a safe, easy to use, and effective means of automating high pressure test and calibration of a wide range of pressure measuring devices. In control mode, these controllers simultaneously measure and control pressure, and can be used for the calibration and testing of pressure gauges, transducers, pressure switches, and production pressure instruments.
The Model 7350 and Model 7615 controllers use a quartz transducer for unmatched performance in conducting automated high pressure testing and calibration. Precision levels are 0.01% of full scale for pressures to 20,000 psi(1400 bar) and 0.02% of full scale for pressures above 20,000 psi (2750 bar). Stability is 0.01% of full scale per year, providing for an annual recalibration interval.
- Pressure ranges up to 10,000 psi (700 bar)
- Precision to 0.01% FS
- Control stability 0.007% FS
- User-friendly menu driven operation
- RS232 & IEEE-488 interface
|
General |
Standard Pressure Ranges (FS, bara) |
6000, 10,000 psia (415, 700 bara) |
Tare Mode |
Tare mode included for gauge calibrations |
Display |
TFT, VGA, active matrix, 6.4 in (162.5 mm) 640 x 480 resolution, 65,000 colors |
Electrical Power |
110 or 220 VAC (±18%), 50/60 Hz, single phase |
Operating Temperature |
41°F to 122°F (5°C to 50°C) |
Storage Temperature |
-4°F to 158°F (-20°C to 70°C) |
Humidity |
5 to 95% relative humidity, non-condensing |
Dimensions (h x w x d) |
17 in x 19 in x 24 in (43 cm x 49 cm x 61 cm) |
Weight |
120 lb (54.5 kg) |
Performance |
Precision |
0.01% of range; to 10,000 psi (700 bar) |
Stability |
0.01% of range, per year |
Control Stability |
0.007% maximum FS for load volume 3 to 35 in3 (7.5 to 89 cm3); control low limit is 10% of maximum FS pressure |
Warm Up Time |
30 minutes, may be left on indefinitely |
Pressure Medium |
Dry, clean air or nitrogen |
Supply Pressure |
10% above FS pressure range (may require optional external gas booster system), 100 to 110 psi (7 to 7.5 bar) industrial grade (shop) air for controller operation |
Pneumatics |
Pneumatics Ports |
Test port: 1/8 in NPT
Supply pressure port: 1/8 in NPT
Air supply: 1/4 in NPT |
Overpressure Protection |
Test port relief valve set at 110% FS
Gas supply port set at 120% FS
Software limits (set by user) |
Communications |
Communications |
RS232 and IEEE-488
Syntax: SCPI
Ruska Series 6000 and Druck DPI 510 emulation |
Drivers |
LabVIEW driver available
MET/CAL® driver available
Intecal software available |
Calibration |
Calibration |
A calibration report providing traceability to the National Institute of Standards and Technology (NIST) is provided with each instrument. NVLAP accredited calibration to ISO 17025 is available as an option.
Recommended calibration interval is one year. |
Options |
Options |
Triple scale
Intecal software |
|
|
|
|
|